Go Direct® Chloride Ion-Selective Electrode BNC

Connect the Go Direct® Chloride Ion-Selective Electrode BNC to our Go Direct® Ion-Selective Electrode Amplifier to measure the concentration of chloride (Cl) ions in aqueous samples.

SKU: GDX-CL-BNC Categories: ,

$458.00

Product Families

The Go Direct® Vernier Chloride Ion-Selective Electrode BNC is a BNC-terminated electrode used to measure the concentration of chloride (Cl) ions in aqueous samples. Requires a Go Direct® Ion-Selective Electrode Amplifier.

Range (mV):Go Direct® Ion-Selective Electrode Amplifier: -1000 mV to +1000 mV

Range (concentration): 1 to 35,000 mg/L (or ppm)
Reproducibility (precision): ±30mV
Interfering ions: CN–, Br–, I–, S2-, OH–, NH3
pH range: 2–12 (no pH compensation)
Temperature range: 0–80°C (no temperature compensation)
Electrode slope: 56 ±3 mV/decade at 25°C
Electrode resistance: 1 to 5 MΩ
Minimum sample size: must be submerged 2.8 cm (1.1 in)

Go Direct® Chloride Ion-Selective Electrode with BNC terminated end, packed with a storage bottle
30 mL bottle of High Standard solution with SDS (100 mg/L Cl–)
30 mL bottle of Low Standard solution with SDS (1 mg/L Cl–)
Short-Term ISE Soaking Bottle

Electrode Storage Bottles (BTL-ES)
Standard Cl ISE Solution ()

User Manual: https://www.vernier.com/manuals/gdx-cl-bnc/
Warranty: 5-year limited warranty (consumables excluded)
FAQ: